Laboratory autoclave LABOKLAV 80-B
Catalog nr.: 210080001000
• Can be used as a stand-alone and tabletop device
• Usable chamber volume 80 l
• Usable space (ØxL in mm) 410 x 610
• Operating pressure 2.8/-1 bar
• Sterilization temperature 98°C - 138°C
• TÜV-certified motorized safety lock
• Separate, integrated steam generator (6,6 kW) with the latest technology and dry heating elements
• Integrated deionized water supply tank for steam generation (automatic and manual filling), level monitoring
• Thermal barrier according to TRB 402 and DIN 61010-2
• Flexible PT 100 reference sensor for liquid sterilization (time triggering via media temperature)
• Gravitational ventilation as well as pressure change ventilation (destruction sterilization)
• Integrated evaporative condensation, water-cooled, temperature-controlled, adjustable
• Validation port
• Safety valve, TÜV-certified component
• Internal memory chip to store up to 150 cycles (can be read out later)
• Collection connection for drainage pipes
• Interior floor support
• Timer function
• Powder-coated casing
• Fully automatic microprocessor control with the following programs: open programs - 3 solid programs - 2 programs for waste sterilization in destruction bags - 1 liquid waste program - 4 liquid programs password-protected programs - 10 freely programmable programs as solid, waste, or liquid programs
Each program can be configured by the operator (sterilization time, sterilization temperature). A custom name can be assigned to each program (per customer request)
Control / Recording
• True 32-bit system with large multicolor touchscreen and 16 GB internal memory
• Additional membrane keyboard with clear and easy-to-understand symbols
• All process data displayed
• Up to 20 programs available at the touch of a button and individually customizable
• Code-protected program access available
• Automatic documentation as a PDF file directly in the internal memory of the controller
• All relevant process data and error messages in a batch protocol
• Clear graphical evaluation
• Data transfer via USB: Automatic copying to a USB stick
• USB port standard
• 16 GB memory chip for archiving at least 100,000 batch protocols standard