Description
Drying and heating oven FD115-230V is a forced-convection laboratory oven built for rapid drying and sterilisation applications. With a nominal volume of 116 l, it covers a temperature range from +10 °C above ambient temperature up to +300 °C and maintains high temperature accuracy via APT.line™ technology. Fully homogeneous temperature distribution, fast thermal dynamics and a powerful fan help reduce process duration when short cycle times matter. An LCD controller, electromechanical exhaust air flap control and forced-air circulation enable reliable regulation of operating conditions. Two chrome-plated racks are included, the chamber is stackable, and an integrated class 2 temperature selector limiter in accordance with DIN 12880 supplies optical alarm protection. Ergonomic handle design and a USB connection for data recording support convenient handling and process documentation. Up to 30 % more energy-efficient than conventional comparable units, this FD Avantgarde.Line oven from manufacturer Binder is a practical choice for laboratory drying, heating and sterilisation work requiring uniform heat and responsive performance.
This article is part of Drying ovens & climate chambers, in Heating, drying & cooling.
Key features
- Pack size: 1 pc(s)
- Nominal volume: 116 l
- Temperature range: +10 °C above ambient to +300 °C
- Max. operating temperature: 300 °C
- Forced air circulation: Yes
- Interface: USB
- Manufacturer: Binder
- UNSPSC: 41104502
- Interior dimensions: 550 × 385 × 550 mm (W × D × H)
- Exterior dimensions: 710 × 605 × 735 mm (W × D × H)
- Weight, empty: 54 kg
- Voltage: 230 V, 50/60 Hz
- Nominal power: 1.3 kW
- Temperature uniformity at 150 °C: ±1.7 K
- Temperature fluctuation at 150 °C: ±0.3 K
- Heating-up time to 150 °C: 18 min
- Recovery time after a 30 s door opening at 150 °C: 4 min
- Air change rate at 100 °C: 32 x/h
- Shelves, standard / maximum: 2/5
- Load per rack: 30 kg
- Permitted load: 150 kg
- Energy consumption at 100 °C: 215 Wh/h
- Energy consumption at 150 °C: 340 Wh/h
- Wall clearance, side: 100 mm
- Wall clearance, back: 160 mm
- Sound pressure level: 43 dB(A)
- Technical Information Sheet:
- Data Sheet:
Recently viewed
You may also like
Questions about Drying and heating chambers FD115-230V
What is this article?
Drying and heating chambers FD115-230V is the article on this page. Luminix Health distributes Binder for laboratory use. It sits in Drying ovens & climate chambers, inside Heating, drying & cooling, in Lab instruments.
A drying oven holds a temperature above ambient to drive off water or to condition a sample. A climate chamber also holds humidity. Uniformity across the chamber, not only the set point, decides whether two samples on different shelves saw the same conditions.
Temperature equipment either holds a set point or drives a sample toward one. An oven dries. A furnace reaches the temperatures of ashing and melting. A bath surrounds a vessel with water or oil. A circulator moves that fluid. An incubator holds the mild warmth of a culture. An autoclave uses steam under pressure to sterilise. The number that matters is the temperature the article actually reaches, not the number painted on the dial. Temperature is part of the international system kept by the BIPM and compared by EURAMET.
Which category should I open next?
The category is Heating, drying & cooling. The sub-category is Drying ovens & climate chambers. The catalogue is Lab instruments.
Shelves beside this one are Muffle & tube furnaces, Heating mantles, Circulators, chillers & thermostats, Water & oil baths, Hot plates & heating blocks, Incubators, Heating & cooling (other), Temperature control accessories, Laboratory refrigerators & freezers, Autoclaves & sterilisers, Bunsen & gas burners and Heating tapes, cables & plates.
What should I check before I order?
When this page shows a safety data sheet, use that manufacturer file. Hazard text does not belong in the sales description. For a pack size, a grade or a certificate that is not on the page, send a quote request or contact the team.
How does ordering and shipping work?
Luminix Health distributes this article from the heart of Europe. Articles in stock ship within 2–5 business days. Standard delivery is free from €250 excl. VAT. Cold-chain goods and dangerous goods under ADR are never included in that free delivery. The terms are in the Shipping Policy. The full brand index is on shop by brand. Who we are is on about us.
Which other categories should I open?
The seven catalogues are Labware, Lab chemicals, Lab instruments, Lab consumables, Chromatography equipment, Lab safety equipment and Life Science & Microbiology.