Description
KBF 720 (E7) (120V) is a large constant climate chamber from the KBF series, engineered for laboratory use over a wide temperature and humidity range. This standard model delivers 727 L interior volume and is specified as KBF720UL-120V for connection to 120 V 1~ 50/60 Hz. The chamber supports controlled environmental conditions for stability testing, bulk sample conditioning, and other climate-dependent laboratory workflows that need consistent temperature and humidity control. Net weight is 344 kg and gross weight 346 kg, reflecting a substantial freestanding instrument for continuous professional operation. Supplied in standard KBF series form by manufacturer Binder, the KBF 720 (E7) (120V) addresses sites that require high chamber capacity within a single constant-climate unit. Pack size is one piece. Origin is Germany (Baden-Württemberg). The appliance is positioned for reproducible climate performance across an extended temperature and humidity envelope as defined for the KBF standard line.
This article is part of Drying ovens & climate chambers, in Heating, drying & cooling.
Key features
- Pack size: 1 pc
- Product type: Constant climate chamber
- Series: KBF (standard)
- Interior volume: 727 L
- Electrical connection: 120 V 1~ 50/60 Hz
- Manufacturer: Binder
- Net weight: 344 kg
- UNSPSC: 41104900
- Interior dimensions: 1011 × 575 × 1250 mm (W × D × H)
- Exterior dimensions: 1285 × 900 × 1951 mm (W × D × H)
- Weight, empty: 298 kg
- Voltage: 120 V, 50/60 Hz
- Nominal power: 1.6 kW
- Temperature range: 0...70 °C
- Temperature range with humidity: 10...70 °C
- Temperature uniformity at 25 °C and 60% RH: ±0.3 K
- Temperature uniformity at 40 °C and 75% RH: ±0.2 K
- Temperature fluctuation at 25 °C and 60% RH: ±0.1 K
- Temperature fluctuation at 40 °C and 75% RH: ±0.1 K
- Recovery time after a 30 s door opening at 25 °C and 60% RH: 2 min
- Recovery time after a 30 s door opening at 40 °C and 75% RH: 5 min
- Humidity range: 10...80 % RH
- Humidity fluctuation at 25 °C and 60% RH: ±1.5 % RH
- Humidity fluctuation at 40 °C and 75% RH: ±1.5 % RH
- Max. heat compensation at 40 °C: 640 W
- Inner doors: 2
- Shelves, standard / maximum: 2/16
- Load per rack: 40 kg
- Permitted load: 200 kg
- Energy consumption at 25 °C and 60% RH: 385 Wh/h
- Energy consumption at 40 °C and 75% RH: 440 Wh/h
- Wall clearance, side: 180 mm
- Wall clearance, back: 100 mm
- Sound pressure level at 25 °C and 60% RH: 51 dB(A)
- Data Sheet:
Recently viewed
You may also like
Questions about KBF 720 (E7) (120V)
What is this article?
KBF 720 (E7) (120V) is the article on this page. Luminix Health distributes Binder for laboratory use. It sits in Drying ovens & climate chambers, inside Heating, drying & cooling, in Lab instruments.
A drying oven holds a temperature above ambient to drive off water or to condition a sample. A climate chamber also holds humidity. Uniformity across the chamber, not only the set point, decides whether two samples on different shelves saw the same conditions.
Temperature equipment either holds a set point or drives a sample toward one. An oven dries. A furnace reaches the temperatures of ashing and melting. A bath surrounds a vessel with water or oil. A circulator moves that fluid. An incubator holds the mild warmth of a culture. An autoclave uses steam under pressure to sterilise. The number that matters is the temperature the article actually reaches, not the number painted on the dial. Temperature is part of the international system kept by the BIPM and compared by EURAMET.
Which category should I open next?
The category is Heating, drying & cooling. The sub-category is Drying ovens & climate chambers. The catalogue is Lab instruments.
Shelves beside this one are Muffle & tube furnaces, Heating mantles, Circulators, chillers & thermostats, Water & oil baths, Hot plates & heating blocks, Incubators, Heating & cooling (other), Temperature control accessories, Laboratory refrigerators & freezers, Autoclaves & sterilisers, Bunsen & gas burners and Heating tapes, cables & plates.
What should I check before I order?
When this page shows a safety data sheet, use that manufacturer file. Hazard text does not belong in the sales description. For a pack size, a grade or a certificate that is not on the page, send a quote request or contact the team.
How does ordering and shipping work?
Luminix Health distributes this article from the heart of Europe. Articles in stock ship within 2–5 business days. Standard delivery is free from €250 excl. VAT. Cold-chain goods and dangerous goods under ADR are never included in that free delivery. The terms are in the Shipping Policy. The full brand index is on shop by brand. Who we are is on about us.
Which other categories should I open?
The seven catalogues are Labware, Lab chemicals, Lab instruments, Lab consumables, Chromatography equipment, Lab safety equipment and Life Science & Microbiology.