The filter shall remove free chlorine to a level below 0.05 mg/l when operated within normal operating specifications. The system provides continuous filtered water through the use of a
duplex (two tank) configuration.
System backwashes are initiated by a water meter. The water meter measures the processed volume and is adjustable.
The backwash control valve is mounted on top (on the media tank) and made from non-corrosive materials.
The control valve has three operating cycles, including operation, backwash and operating flow rinse.
The tanks are made of glass fiber reinforced polypropylene, are designed for a maximum operating pressure of 8.6 bar (8.8 kg/cm2) and have been hydrostatically tested at 20.7 bar.
Each system includes acid-washed carbon media. The media must be solid and have a suitable particle size of 20-40 mesh.
Technical specifications:
Flow Range: 30-57 lpm 8-15 gpm
Optimal Media Flow Rate: 15 lpm (0,9m³/h) 4.0 gpm
Pressure Range: 1 – 8.6 bar 12 – 125 PSI
Temperature Range: 1 – 49°C 35 - 120°F
Media Vessel Construtcion: PE wrapped Polyethylene
Media Volume: (2x) 19.8 l
Regeneration Control: non-electric Use Meter
Metering Flow Range: 01.1-94.6 lpm 0.30 – 25.0 gpm
Service: Downflow
Backwash: Upflow
Automatic switching to a second media vessel, therefore operation is also possible during backwashing!
Connections:
Inlet, Outlet: 1'' Male Thread
Drain: 0,5'' Tube
Auxiliary Drain Line: 0,5'' Tube
Meter Disc Selection:
• for setting the backwash intervals
• Disc No. 4 is pre-installed
• All discs are included in delivery
Recently viewed
You may also like
Browse related categories
Category
Sub-category
Water purification cartridges & consumables
Catalogue
Brand
Questions about ACTIVE CARBON, Dual Tank Backwash filter, 15LPM
What is this article?
ACTIVE CARBON, Dual Tank Backwash filter, 15LPM is the article on this page. Luminix Health distributes Evoqua Water Technologies for laboratory use. It sits in Water purification cartridges & consumables, inside Water purification, in Instrumenten.
Cartridges, lamps and final filters are the consumables of a water system. A spent ion-exchange cartridge lets ions through, and the resistivity falls. The cartridge has to be the one that model names. A cartridge from a different brand with a similar shape is a leak or a bypass.
Laboratory water is a reagent. Tap water carries ions, organics and microbes that become part of every result they touch. A pure-water system takes most of them out. An ultrapure system, often a polisher after a first stage, is what HPLC, cell culture and trace analysis ask for. Resistivity, total organic carbon and endotoxin are different claims. A cartridge is spent when the number drifts, not when the calendar says so. The measurement of those impurities sits with NIST and the BIPM.
Which category should I open next?
The category is Water purification. The sub-category is Water purification cartridges & consumables. The catalogue is Instrumenten.
Shelves beside this one are Ultrapure & pure water systems.
What should I check before I order?
When this page shows a safety data sheet, use that manufacturer file. Hazard text does not belong in the sales description. For a pack size, a grade or a certificate that is not on the page, send a quote request or contact the team.
How does ordering and shipping work?
Luminix Health distributes this article from the heart of Europe. Articles in stock ship within 2–5 business days. Standard delivery is free from €250 excl. VAT. Cold-chain goods and dangerous goods under ADR are never included in that free delivery. The terms are in the Shipping Policy. The full brand index is on shop by brand. Who we are is on about us.
Which other categories should I open?
The seven catalogues are Labware, Lab chemicals, Lab instruments, Lab consumables, Chromatography equipment, Lab safety equipment and Life Science & Microbiology.