Description
Pure water system, Smart2Pure UV/UF 6 l/h, incl. Smart2Pure unit with integral tank incl. pressure and recirculation pump, PVC-connecting hose 1.5 m straight/angle, sterile filter capsule 0.2 µm, pretreatment pack and pure water cartridge, PE-connecting hose 6/4
Compact system for the production of contamination-free water from mains water. The device is connected directly to the mains water supply.
- ASTM Type I (resistance of 18.2 MΩ x cm)
- ASTM Type II (resistance of 10-15 MΩ x cm)
- Combined cartridge (reverse osmosis membrane/activated charcoal) for pre-treatment of feed water
- Direct tapping of type I ultrapure water through a sterile end filter (0.2 µm)
- Integrated 6-litre tank for storing type II pure water
- Guaranteed pure water quality: the recirculation pump and sterile air filter protect from microbial contamination
- Additional pressure outlet on ultrapure water tank for supplying downstream devices
- Simple and economical change of expendables as treatment cartridges can be replaced separately
- Space-saving mounting on laboratory bench or on the wall
| Drinking water requirements: | |
| Drinking water to DIN 2000 | Manganese and iron content: <0.05 mg/l |
| Chlorine concentration: <0.1 mg/l | |
| Colloid index: <3 | |
| Feed water temperature: 2-35 °C | |
| Feed water pressure: 1-6 bar | |
With UV lamp and UF module. The UF ultra-filtration module guarantees almost complete elimination of macro-molecular contamination such as micro-organisms, pyrogens (bacterial endotoxins), nucleases and particles; the combination of UV oxidation (254 and 185 nm) and ultra-filtration produces analytically pure water with a specific resistance of up to 18.2 MΩ cm, TOC values of <5 ppb, pyrogene levels of <0.001 EU/ml and RNase, DNase and DNA concentrations below the detection limit.Typical applications: molecular biology, PCR, DNA, monoclonal antibodies, cell culture media.
| Type | Smart2Pure UV/UF 6 l/h |
| Conductivity of high-purity water, type I | 0.055 µS/cm |
| Conductivity of pure water, type II | 0,067-0,1 µS/cm |
| Resistance of high-purity water, type I | 18.2 MΩ·cm |
| Resistance of pure water, type II | 10-15 MΩ·cm |
| Withdrawal capacity of high-purity water, type I | 0.6 l/min |
| Withdrawal capacity of pure water, type II | 6 l/h |
| Bacteria content of high-purity water, type I | <1 KbE/ml |
| Endotoxin content of high-purity water, type I | 0.001 EU/ml |
| TOC value of high-purity water, type I | 1-5 ppb |
| Particle content of high-purity water, type I | <1 particle per ml (particle size <0.2 μm) |
| Width | 305 mm |
| Depth | 400 mm |
| Height | 545 mm |
| Weight | 22 kg |
| Connection | Feedwater: hose (length 1.5 m) with Ø outside 8 mm and 3/4″ Waste water: 1/4″ or Ø outside 8 mm, hose (length 2 m), pressure-free discharge |
| Mains connection | 100-240 V, 50/60 Hz |