Description
Pure water production from softened potable tap water. The system combines the technology of the reverse osmosis with the final deionization of the permeate by single bed electroeioni-
zation. The system is placed in a well-designed plastic housing.
The reverse osmosis component mainly consists of:
1 Solenoid valve for raw water
1 Pre-treatment module AMD, fine filter made of plastic, inclusive active carbon/fine filter cartridge 5 µm, front-side operated
1 Pressure switch as pump protection
1 High pressure pump made of brass with magnetic-coupled E-motor
1 Pressure gauge to indicate operating pressure
1 Module set with spiral-wound membranes, made of polyamide on a polysulphone basis, incl. pressure vessel with end caps
1 Solenoid valve for permeate reject when conductivity is exceeded
1 Conductivity meter for permeate
1 Restrictor for limiting the concentrate quantity
The Ionpure component mainly consists of:
1 Ionpure cell, two-pass, with chambers for anion and cation exchange resin and brine chambers, anodes, cathodes, ion-selective membranes and end plates
1 Conductivity meter for pure water
1 Restrictor for limiting the brine
1 Conditioning cartridge
Control:
Display with membrane keyboard with monitor for:
- Conductivity permeate and pure water
- Level indication storage tank in Percent for level sensors or fixed value for float switches
- 3 free selectable switching activating points ON/OFF and running dry-protection for booster pump
- Status indication via symbols
- Indications for maintenance intervals and replacement of cartridge
- Fault indication (insufficient raw water, insufficient permeate, conductance LF1 / LF 2)
- Timer program for displacement of concentrate and module flushing
- RS 232 interface (processing software optional)
- Multilingualism
1 set connecting hoses
1 filter cup wrench
1 operator's manual
Performance specifications:
- Production rate at 15°C: 30 l/h
- Conductivity: < 0.2 µS/cm
- typical conductivity: 0.07 µS/cm
- Silicate content: < 10 ppb
- TOC value: < 30 ppb
- particle rejection: > 99 %
Feed water specifications:
- Flow rate: 44 - 66 l/h
- Pressure: 0,1 - 5 bar
- Maximum conductivity: < 1400 µS/cm
- SDI: < 3
- FE and free chlorine: < 0,1 mg/l
- CO2 value in the feed: < 10 mg/l
- max. SiO2 value in the feed: < 10 mg/l
- Permissible water temp.: 5 - 30°C
Technical data:
- Operating pressure: max. 6 bar
- Electrical power: 100-240V, 50-60 Hz
- Current consumption: 0.2 kW
- Fuse: 16 A
- Permissible ambient temp.: 5 - 35°C
Connections:
Raw water: 3/4''
Pure water: d= 8 mm
Drain concentrate: d= 8 mm
Drain brine: d= 6 mm
Dimensions and Weight:
Height: 530 mm
Width: 340 mm
Depth: 420 mm
Weight: 35 kg
Please note:
The performance is based on a raw water salt content of max. 750 mg/l (TDS) equivalent to a conductivity of 1400 µS/cm at 15°C. From 10°dH or higher, a softener is recommended.
All a. m. components are assembled into a complete system. The Ultra Clear RO EDI can be put into operation without the help of a factory service engineer.
The feed water specifications are guidelines. The pure water specifications can only be guaranteed, if a drinking water analysis is available for the plant location and all other relevant
parameters have been evaluated.
Key features
Récemment consulté
Vous aimerez peut-être aussi
Questions about Ultra Clear RO EDI 30 Reverse osmosis/ EDI combination Pure Water System
What is this article?
Ultra Clear RO EDI 30 Reverse osmosis/ EDI combination Pure Water System is the article on this page. Luminix Health distributes Evoqua Water Technologies for laboratory use. It sits in Ultrapure & pure water systems, inside Water purification, in Instruments de précision.
A pure-water system, often by reverse osmosis and deionisation, feeds the laboratory. An ultrapure polisher takes that water to the resistivity and organic levels HPLC or cell culture ask for. Production rate and the feed water it expects are the specification. The system is not a cooler.
Laboratory water is a reagent. Tap water carries ions, organics and microbes that become part of every result they touch. A pure-water system takes most of them out. An ultrapure system, often a polisher after a first stage, is what HPLC, cell culture and trace analysis ask for. Resistivity, total organic carbon and endotoxin are different claims. A cartridge is spent when the number drifts, not when the calendar says so. The measurement of those impurities sits with NIST and the BIPM.
Which category should I open next?
The category is Water purification. The sub-category is Ultrapure & pure water systems. The catalogue is Instruments de précision.
Shelves beside this one are Water purification cartridges & consumables.
What should I check before I order?
When this page shows a safety data sheet, use that manufacturer file. Hazard text does not belong in the sales description. For a pack size, a grade or a certificate that is not on the page, send a quote request or contact the team.
How does ordering and shipping work?
Luminix Health distributes this article from the heart of Europe. Articles in stock ship within 2–5 business days. Standard delivery is free from €250 excl. VAT. Cold-chain goods and dangerous goods under ADR are never included in that free delivery. The terms are in the Shipping Policy. The full brand index is on shop by brand. Who we are is on about us.
Which other categories should I open?
The seven catalogues are Labware, Lab chemicals, Lab instruments, Lab consumables, Chromatography equipment, Lab safety equipment and Life Science & Microbiology.