Description
LAUDA PRO RP 290 EC
Cooling heating circulator 230 V; 50 Hz
Features:
Thermostatic circulator with latest microprocessor technology. Integrated cooling system
Large 5,7'' Multi-Touch-Color display
Operation with multi touch display
Command console can be detached and used as remote control
User administration for up to 20 user
Data in- and export via USB
Timer und count-down-function
EasyUse system for simple operation of the whole unit
SelfCheck Assistant for system diagnosis
USB and Ethernet interface equipped as standard
Retrofittable for upgrading with interface (RS 232/485, Profibus; EtherCAT; analogue or contact modules)
Fully electronic continuous controller with PID action for internal & external control
PowerAdapt system for the use of the maximum possible amount of heat permitted by the power supply system
Low-level and adjustable over-temperature protection with acoustic alarm for use with flammable and non-flammable liquids
Programmer with up to 100 programs, up to 50 segments each
Tower design with small footprint
LAUDA Varioflex pump (pressure/suction) with 8 selectable levels, connections rear
Low filling volume enables rapid changes in temperature
Bath vessel made of stainless steel (insulated, with overflow and drain tap)
SmartCool system for energy-saving digital cooling management including compressor on-off control
Hybrid cooling system: (heat removal complementary with cooling water)
Utilizes natural refrigerants
Technical Features (according to DIN 12876):
Working temperature range: -90 ... 200 °C
Operating temperature range: -90 ... 200 °C
Ambient temperature range: 5 ... 40 °C
Temperature stability: 0.05 ± K
Heater power max.: 2.5 kW
Power consumption max.: 3.7 kW
Current max.: 16 A
Pump Pressure max.: 0,7 bar
Pump suction max.: 0,4 bar
Pump flow rate max. (pressure): 22 L/min
Pump Flow rate max. (suction): 20 L/min
In / Outlet connection thread (outside): M16 x 1
In / Outlet Ø olives: 13 mm
Filling volume min.: 2.4 L
Filling volume max.: 4.4 L
Overall dimensions (WxDxH): 390 x 600 x 685 mm
Weight: 79 kg
Refrigerant stage 1: R-290 (GWP 3); 0.120 kg; 0.0 t CO2-eq
Refrigerant stage 2: R-170 (GWP 6); 0.070 kg; 0.0 t CO2-eq
Power supply: 230 V; 50 Hz
Power plug: Power cord with angled plug (CEE7/7)
Cooling Capacity (Pumpenstufe):
20 °C (8); Ethanol; 0,8 kW
10 °C (8); Ethanol; 0,77 kW
0 °C (8); Ethanol; 0,74 kW
-10 °C (8); Ethanol; 0,72 kW
-20 °C (4); Ethanol; 0,7 kW
-30 °C (4); Ethanol; 0,68 kW
-40 °C (4); Ethanol; 0,64 kW
-50 °C (4); Ethanol; 0,56 kW
-60 °C (4); Ethanol; 0,39 kW
-70 °C (4); Ethanol; 0,21 kW
-80 °C (4); Ethanol; 0,09 kW
-90 °C (2); Ethanol; 0,01 kW
Standard Accessory:
2 screw caps, 2 closing plugs
2 nipples 13 mm for pump connectors
2 nipples 10 mm with screw cap G3/8 for cooling water
Cooling heating circulator 230 V; 50 Hz
Features:
Thermostatic circulator with latest microprocessor technology. Integrated cooling system
Large 5,7'' Multi-Touch-Color display
Operation with multi touch display
Command console can be detached and used as remote control
User administration for up to 20 user
Data in- and export via USB
Timer und count-down-function
EasyUse system for simple operation of the whole unit
SelfCheck Assistant for system diagnosis
USB and Ethernet interface equipped as standard
Retrofittable for upgrading with interface (RS 232/485, Profibus; EtherCAT; analogue or contact modules)
Fully electronic continuous controller with PID action for internal & external control
PowerAdapt system for the use of the maximum possible amount of heat permitted by the power supply system
Low-level and adjustable over-temperature protection with acoustic alarm for use with flammable and non-flammable liquids
Programmer with up to 100 programs, up to 50 segments each
Tower design with small footprint
LAUDA Varioflex pump (pressure/suction) with 8 selectable levels, connections rear
Low filling volume enables rapid changes in temperature
Bath vessel made of stainless steel (insulated, with overflow and drain tap)
SmartCool system for energy-saving digital cooling management including compressor on-off control
Hybrid cooling system: (heat removal complementary with cooling water)
Utilizes natural refrigerants
Technical Features (according to DIN 12876):
Working temperature range: -90 ... 200 °C
Operating temperature range: -90 ... 200 °C
Ambient temperature range: 5 ... 40 °C
Temperature stability: 0.05 ± K
Heater power max.: 2.5 kW
Power consumption max.: 3.7 kW
Current max.: 16 A
Pump Pressure max.: 0,7 bar
Pump suction max.: 0,4 bar
Pump flow rate max. (pressure): 22 L/min
Pump Flow rate max. (suction): 20 L/min
In / Outlet connection thread (outside): M16 x 1
In / Outlet Ø olives: 13 mm
Filling volume min.: 2.4 L
Filling volume max.: 4.4 L
Overall dimensions (WxDxH): 390 x 600 x 685 mm
Weight: 79 kg
Refrigerant stage 1: R-290 (GWP 3); 0.120 kg; 0.0 t CO2-eq
Refrigerant stage 2: R-170 (GWP 6); 0.070 kg; 0.0 t CO2-eq
Power supply: 230 V; 50 Hz
Power plug: Power cord with angled plug (CEE7/7)
Cooling Capacity (Pumpenstufe):
20 °C (8); Ethanol; 0,8 kW
10 °C (8); Ethanol; 0,77 kW
0 °C (8); Ethanol; 0,74 kW
-10 °C (8); Ethanol; 0,72 kW
-20 °C (4); Ethanol; 0,7 kW
-30 °C (4); Ethanol; 0,68 kW
-40 °C (4); Ethanol; 0,64 kW
-50 °C (4); Ethanol; 0,56 kW
-60 °C (4); Ethanol; 0,39 kW
-70 °C (4); Ethanol; 0,21 kW
-80 °C (4); Ethanol; 0,09 kW
-90 °C (2); Ethanol; 0,01 kW
Standard Accessory:
2 screw caps, 2 closing plugs
2 nipples 13 mm for pump connectors
2 nipples 10 mm with screw cap G3/8 for cooling water