Description
Dynamic climate chamber MKF1020-400V is an environmental simulation chamber for cold and heat testing according to common temperature and climate test standards under DIN and IEC norms. With a nominal volume of 991 l it operates from −40 °C to +180 °C and regulates relative humidity from 10 % r.H. to 98 % r.H. via a capacitive humidity sensor with steam humidification. APT.line™ preheating chamber technology supports uniform conditions, while an intuitive touchscreen controller provides time-segment and real-time programming. The chamber from manufacturer Binder uses CFC-free refrigerant R-452A and includes an integrated 20 L water storage tank, one stainless-steel rack, two 80 mm access ports with silicone plugs on the left and right, and four potential-free switching contacts. Program memory, USB interface and timer functions are provided as standard. Rated power consumption is 12500 W. Extensive series equipment and straightforward operation make the unit suitable for demanding climate and temperature tests in laboratory and industrial environments.
This article is part of Drying ovens & climate chambers, in Heating, drying & cooling.
Key features
- Pack size: 1 pc
- Nominal volume: 991 l
- Temperature range: −40 °C to +180 °C
- Humidity range: 10 % r.H. to 98 % r.H.
- Power consumption: 12500 W
- Interface: USB
- Manufacturer: Binder
- UNSPSC: 41104900
- Interior dimensions: 1200 × 810 × 1020 mm (W × D × H)
- Exterior dimensions: 1580 × 1480 × 2005 mm (W × D × H)
- Weight, empty: 636 kg
- Voltage: 400 V, 50 Hz
- Nominal power: 12.5 kW
- Temperature uniformity, depending on setpoint: ±0.1...±1.8 K
- Temperature fluctuation, depending on setpoint: ±0.1...±0.5 K
- Temperature fluctuation with humidity, depending on setpoint: ±0.1...±1.5 K
- Average heating-up rate between -40 °C and 180 °C: 5.5 K/min
- Cooling-down time from 180 °C to -40 °C: 120 min
- Average cooling-down time between 180 °C and -40 °C: 5 K/min
- Max. heat compensation at 25 °C and 90% RH: 1500 W
- Shelves, standard / maximum: 1/11
- Load per rack: 40 kg
- Permitted load: 200 kg
- Energy consumption at 25 °C and 60% RH: 2200 Wh/h
- Wall clearance, side: 200 mm
- Wall clearance, back: 300 mm
- Sound pressure level: 69 dB(A)
- Technical Information Sheet:
- Data Sheet:
Recently viewed
You may also like
Questions about Dynamic climate chambers MKF1020-400V
What is this article?
Dynamic climate chambers MKF1020-400V is the article on this page. Luminix Health distributes Binder for laboratory use. It sits in Drying ovens & climate chambers, inside Heating, drying & cooling, in Lab instruments.
A drying oven holds a temperature above ambient to drive off water or to condition a sample. A climate chamber also holds humidity. Uniformity across the chamber, not only the set point, decides whether two samples on different shelves saw the same conditions.
Temperature equipment either holds a set point or drives a sample toward one. An oven dries. A furnace reaches the temperatures of ashing and melting. A bath surrounds a vessel with water or oil. A circulator moves that fluid. An incubator holds the mild warmth of a culture. An autoclave uses steam under pressure to sterilise. The number that matters is the temperature the article actually reaches, not the number painted on the dial. Temperature is part of the international system kept by the BIPM and compared by EURAMET.
Which category should I open next?
The category is Heating, drying & cooling. The sub-category is Drying ovens & climate chambers. The catalogue is Lab instruments.
Shelves beside this one are Muffle & tube furnaces, Heating mantles, Circulators, chillers & thermostats, Water & oil baths, Hot plates & heating blocks, Incubators, Heating & cooling (other), Temperature control accessories, Laboratory refrigerators & freezers, Autoclaves & sterilisers, Bunsen & gas burners and Heating tapes, cables & plates.
What should I check before I order?
When this page shows a safety data sheet, use that manufacturer file. Hazard text does not belong in the sales description. For a pack size, a grade or a certificate that is not on the page, send a quote request or contact the team.
How does ordering and shipping work?
Luminix Health distributes this article from the heart of Europe. Articles in stock ship within 2–5 business days. Standard delivery is free from €250 excl. VAT. Cold-chain goods and dangerous goods under ADR are never included in that free delivery. The terms are in the Shipping Policy. The full brand index is on shop by brand. Who we are is on about us.
Which other categories should I open?
The seven catalogues are Labware, Lab chemicals, Lab instruments, Lab consumables, Chromatography equipment, Lab safety equipment and Life Science & Microbiology.