ICO105 Plus CO2 incubator, 107 l incl. 2 st. steel perforated sheet, active humidification and dehumidification control and a factory calibration certificate at +37 °C, 5% CO2.
For cell cultures.
- Operating temperature range from +5 °C above room temperature to 50 °C (in 0.1 °C increments) and CO2 concentrations up to 20% (in 0.1% increments)
- Stainless steel interior (1.4301)
- Heated inner glass door with opening (8 mm Ø) for gas sampling
- Stored sterilisation program: 60 minutes at 180 °C
- To supply the incubator with CO2, a separate gas cylinder or a central compressed-gas tank is required
- Demineralised/fully desalinated chlorine-free water with a conductivity of 5 to 10 µS/cm and pH value 5-7 is required for humidification
- With factory calibration certificate at +37 °C, 5% CO2
- With fan (non-adjustable)
- German, English, French and Spanish language options
- Digital timer (adjustable from 1 min to 99 d, 23 h) or continuous operation
- Two graphical TFT displays with touchscreen for operation and displaying temperature, program runtime and graphical display of the target and actual values
- Integrated data logger with 10-year memory
- Display of recorded log data (max. 10,000 values correspond to approx. one week) directly on the device
- With Ethernet communication port for reading out log data as well as for transmitting and launching programs and online logging
- USB port for installing programs and reading out the data logger incl. software (to create ramp programs and weekly programs)
- Definition of individual min./max. values for protection against excessive/insufficient temperatures
Active humidification and dehumidification control microprocessor (40 to 97% RH in 0.5% RH increments) with digital display to prevent condensation formation and enable the target humidity to be reached even more quickly with short recovery times. Humidity supply with water (from external 2.5 l container) via self-priming pump.
| Type | ICO105 Plus |
| Vol. | 107 l |
| Operating temperature | from +5 °C above room temperature to +50 °C |
| Inner width | 560 mm |
| Inner depth | 365 mm |
| Inner height | 480 mm |
| Outer width | 719 mm |
| Outer depth | 647 mm |
| Outer height | 850 mm |
| Required input pressure | 1.0 to 1.2 bar, purity 99.995 vol% |
| Max. removable shelves | 6 |
| Rated power | 1300 W |
| Temporal temp. variance | ±0,1 °C |
| Spatial temp. deviation | ±0,3 °C |
| CO2 constancy | ±0,20 % |
| Weight | 100 kg |
| Mains connection | 230 V, 50/60 Hz |
Recently viewed
You may also like
Browse related categories
Category
Sub-category
Catalogue
Brand
Questions about ICO105 Plus CO2 incubator, 107 l incl. 2 st. steel perforated sheet, 1 unit(s)
What is this article?
ICO105 Plus CO2 incubator, 107 l incl. 2 st. steel perforated sheet, 1 unit(s) is the article on this page. Luminix Health distributes Memmert for laboratory use. It sits in Incubators, inside Heating, drying & cooling, in Lab instruments.
An incubator holds the mild, even temperature a culture needs, often 25 °C or 37 °C, sometimes with carbon dioxide. It is not a drying oven. Uniformity and, for cells, the gas control are the specification. A flask of cells in a drying oven is a dead flask.
Temperature equipment either holds a set point or drives a sample toward one. An oven dries. A furnace reaches the temperatures of ashing and melting. A bath surrounds a vessel with water or oil. A circulator moves that fluid. An incubator holds the mild warmth of a culture. An autoclave uses steam under pressure to sterilise. The number that matters is the temperature the article actually reaches, not the number painted on the dial. Temperature is part of the international system kept by the BIPM and compared by EURAMET.
Which category should I open next?
The category is Heating, drying & cooling. The sub-category is Incubators. The catalogue is Lab instruments.
Shelves beside this one are Muffle & tube furnaces, Heating mantles, Circulators, chillers & thermostats, Drying ovens & climate chambers, Water & oil baths, Hot plates & heating blocks, Heating & cooling (other), Temperature control accessories, Laboratory refrigerators & freezers, Autoclaves & sterilisers, Bunsen & gas burners and Heating tapes, cables & plates.
What should I check before I order?
The technical information on this page is the manufacturer's sheet. Read that, not a rewritten specification.
When this page shows a safety data sheet, use that manufacturer file. Hazard text does not belong in the sales description. For a pack size, a grade or a certificate that is not on the page, send a quote request or contact the team.
How does ordering and shipping work?
Luminix Health distributes this article from the heart of Europe. Articles in stock ship within 2–5 business days. Standard delivery is free from €250 excl. VAT. Cold-chain goods and dangerous goods under ADR are never included in that free delivery. The terms are in the Shipping Policy. The full brand index is on shop by brand. Who we are is on about us.
Which other categories should I open?
The seven catalogues are Labware, Lab chemicals, Lab instruments, Lab consumables, Chromatography equipment, Lab safety equipment and Life Science & Microbiology.