KBF 1600 (E7) (120V) is a constant climate chamber from the KBF series, offering a wide temperature and humidity range and a large interior volume of 1610 L. Configured as the standard model for 120 V 1~ 50/60 Hz operation, it supports professional laboratory work that needs controlled climate conditions over an extensive usable space. The unit is supplied by manufacturer Binder within the KBF constant climate cabinet portfolio and is delivered as one complete piece. Net weight is 491 kg and gross weight is 493 kg, consistent with a high-capacity climate system intended for exclusive non-household laboratory environments. This class of chamber is applied when laboratories require ample volume for conditioning, storage, or testing under defined temperature and humidity. Made in Germany, the KBF 1600 (E7) (120V) pairs the broad climatic operating window of the KBF series with 1610 L capacity, serving demanding laboratory material workflows that depend on reproducible environmental control at large scale.
This article is part of Drying ovens & climate chambers, in Heating, drying & cooling.
Key features
- Pack size: 1 pc(s)
- Product type: Constant climate chamber
- Series: KBF
- Interior volume: 1610 L
- Power supply: 120 V 1~ 50/60 Hz
- Manufacturer: Binder
- Net weight: 491 kg
- UNSPSC: 41104900
- Interior dimensions: 1521 × 847 × 1250 mm (W × D × H)
- Exterior dimensions: 1795 × 1175 × 1951 mm (W × D × H)
- Weight, empty: 438 kg
- Voltage: 120 V, 50/60 Hz
- Nominal power: 1.7 kW
- Temperature range: 0...70 °C
- Temperature range with humidity: 10...70 °C
- Temperature uniformity at 25 °C and 60% RH: ±0.4 K
- Temperature uniformity at 40 °C and 75% RH: ±0.3 K
- Temperature fluctuation at 25 °C and 60% RH: ±0.1 K
- Temperature fluctuation at 40 °C and 75% RH: ±0.1 K
- Recovery time after a 30 s door opening at 25 °C and 60% RH: 3 min
- Recovery time after a 30 s door opening at 40 °C and 75% RH: 7 min
- Humidity range: 10...80 % RH
- Humidity fluctuation at 25 °C and 60% RH: ±1.5 % RH
- Humidity fluctuation at 40 °C and 75% RH: ±1.5 % RH
- Max. heat compensation at 40 °C: 640 W
- Inner doors: 2
- Shelves, standard / maximum: 2/16
- Load per rack: 40 kg
- Permitted load: 200 kg
- Energy consumption at 25 °C and 60% RH: 410 Wh/h
- Energy consumption at 40 °C and 75% RH: 570 Wh/h
- Wall clearance, side: 180 mm
- Wall clearance, back: 100 mm
- Sound pressure level at 25 °C and 60% RH: 54 dB(A)
- Data Sheet:
Recently viewed
You may also like
Questions about KBF 1600 (E7) (120V)
What is this article?
KBF 1600 (E7) (120V) is the article on this page. Luminix Health distributes Binder for laboratory use. It sits in Drying ovens & climate chambers, inside Heating, drying & cooling, in Lab instruments.
A drying oven holds a temperature above ambient to drive off water or to condition a sample. A climate chamber also holds humidity. Uniformity across the chamber, not only the set point, decides whether two samples on different shelves saw the same conditions.
Temperature equipment either holds a set point or drives a sample toward one. An oven dries. A furnace reaches the temperatures of ashing and melting. A bath surrounds a vessel with water or oil. A circulator moves that fluid. An incubator holds the mild warmth of a culture. An autoclave uses steam under pressure to sterilise. The number that matters is the temperature the article actually reaches, not the number painted on the dial. Temperature is part of the international system kept by the BIPM and compared by EURAMET.
Which category should I open next?
The category is Heating, drying & cooling. The sub-category is Drying ovens & climate chambers. The catalogue is Lab instruments.
Shelves beside this one are Muffle & tube furnaces, Heating mantles, Circulators, chillers & thermostats, Water & oil baths, Hot plates & heating blocks, Incubators, Heating & cooling (other), Temperature control accessories, Laboratory refrigerators & freezers, Autoclaves & sterilisers, Bunsen & gas burners and Heating tapes, cables & plates.
What should I check before I order?
When this page shows a safety data sheet, use that manufacturer file. Hazard text does not belong in the sales description. For a pack size, a grade or a certificate that is not on the page, send a quote request or contact the team.
How does ordering and shipping work?
Luminix Health distributes this article from the heart of Europe. Articles in stock ship within 2–5 business days. Standard delivery is free from €250 excl. VAT. Cold-chain goods and dangerous goods under ADR are never included in that free delivery. The terms are in the Shipping Policy. The full brand index is on shop by brand. Who we are is on about us.
Which other categories should I open?
The seven catalogues are Labware, Lab chemicals, Lab instruments, Lab consumables, Chromatography equipment, Lab safety equipment and Life Science & Microbiology.