Description
Drying and heating oven ED 23UL-120V is a compact laboratory unit designed for routine drying and sterilisation work up to +300 °C. With a nominal volume of 20 l and natural convection, thermal processes run efficiently and deliver fast, uniform drying results. Temperature control spans from +5 °C above ambient temperature to +300 °C, supported by APT.line™ preheating chamber technology for homogeneous heat distribution. An adjustable exhaust air flap and a controller with time functions provide flexible process handling, while two chrome-plated racks offer practical loading space. Safety is reinforced by a Class 2 temperature selector limiter according to DIN 12880 with optical alarm. The unit features an Ethernet interface and a United States Type A/B mains plug. Net weight is 28 kg (gross 30 kg). This ED-series oven from manufacturer Binder is well suited to everyday drying and heating tasks in research, quality control and production laboratories where reliable natural-convection performance is required.
This article is part of Drying ovens & climate chambers, in Heating, drying & cooling.
Key features
- Pack size: 1 pc(s)
- Nominal volume: 20 l
- Temperature range: +5 °C above ambient to +300 °C
- Convection: natural convection
- Racks included: 2 chrome-plated shelves
- Interface: Ethernet
- Manufacturer: Binder
- UNSPSC: 41104502
- Interior dimensions: 222 × 300 × 330 mm (W × D × H)
- Exterior dimensions: 435 × 520 × 495 mm (W × D × H)
- Weight, empty: 27 kg
- Voltage: 120 V, 60 Hz
- Nominal power: 0.8 kW
- Temperature uniformity at 150 °C: ±2.8 K
- Temperature fluctuation at 150 °C: ±0.5 K
- Heating-up time to 150 °C: 28 min
- Recovery time after a 30 s door opening at 150 °C: 5 min
- Air change rate at 150 °C: 13 x/h
- Shelves, standard / maximum: 2/4
- Load per rack: 12 kg
- Permitted load: 25 kg
- Energy consumption at 150 °C: 148 Wh/h
- Wall clearance, side: 50 mm
- Wall clearance, back: 100 mm
- Technical Information Sheet:
- Data Sheet:
Recently viewed
You may also like
Questions about Drying and heating chambers ED 23UL-120V
What is this article?
Drying and heating chambers ED 23UL-120V is the article on this page. Luminix Health distributes Binder for laboratory use. It sits in Drying ovens & climate chambers, inside Heating, drying & cooling, in Lab instruments.
A drying oven holds a temperature above ambient to drive off water or to condition a sample. A climate chamber also holds humidity. Uniformity across the chamber, not only the set point, decides whether two samples on different shelves saw the same conditions.
Temperature equipment either holds a set point or drives a sample toward one. An oven dries. A furnace reaches the temperatures of ashing and melting. A bath surrounds a vessel with water or oil. A circulator moves that fluid. An incubator holds the mild warmth of a culture. An autoclave uses steam under pressure to sterilise. The number that matters is the temperature the article actually reaches, not the number painted on the dial. Temperature is part of the international system kept by the BIPM and compared by EURAMET.
Which category should I open next?
The category is Heating, drying & cooling. The sub-category is Drying ovens & climate chambers. The catalogue is Lab instruments.
Shelves beside this one are Muffle & tube furnaces, Heating mantles, Circulators, chillers & thermostats, Water & oil baths, Hot plates & heating blocks, Incubators, Heating & cooling (other), Temperature control accessories, Laboratory refrigerators & freezers, Autoclaves & sterilisers, Bunsen & gas burners and Heating tapes, cables & plates.
What should I check before I order?
When this page shows a safety data sheet, use that manufacturer file. Hazard text does not belong in the sales description. For a pack size, a grade or a certificate that is not on the page, send a quote request or contact the team.
How does ordering and shipping work?
Luminix Health distributes this article from the heart of Europe. Articles in stock ship within 2–5 business days. Standard delivery is free from €250 excl. VAT. Cold-chain goods and dangerous goods under ADR are never included in that free delivery. The terms are in the Shipping Policy. The full brand index is on shop by brand. Who we are is on about us.
Which other categories should I open?
The seven catalogues are Labware, Lab chemicals, Lab instruments, Lab consumables, Chromatography equipment, Lab safety equipment and Life Science & Microbiology.