Description
Dynamic climate chamber MKFT720-400V is a large-capacity environmental simulation cabinet specialised for dynamic climate changes between −70 °C and +180 °C with controllable humidity from 10 % r.H. to 98 % r.H. Its 728 l stainless-steel interior, APT.line™ preheating chamber technology and cascade compression cooling with CFC-free refrigerants R-452A and R-23 provide high performance reserves and fast pull-down for complex, standards-compliant climate tests. Humidity is regulated by a capacitive sensor with steam humidification; a 20 L integrated water reservoir, heated viewing window with LED lighting and programmable condensation protection protect samples during operation. An intuitive touchscreen controller supports time-segment and real-time programming, an internal data logger allows USB readout of open-format measured values, and BINDER Multi Management Software APT-COM™ Basic Edition is included. Two 80 mm access ports with silicone plugs, four potential-free switching contacts, class 2 temperature safety limiter (DIN 12880), door heating, adjustable ramp function, integrated chart recorder and Ethernet interface round out the feature set. Mounted on four casters (two with brakes), this unit from manufacturer Binder is designed for high-class cyclic climate testing in demanding laboratory environments.
This article is part of Drying ovens & climate chambers, in Heating, drying & cooling.
Key features
- Pack size: 1 pc
- Nominal volume: 728 l
- Temperature range: −70 °C to +180 °C
- Humidity range: 10 % r.H. to 98 % r.H.
- Power consumption: 11000 W
- Access ports: 2 × 80 mm with silicone plugs
- Manufacturer: Binder
- UNSPSC: 41104900
- Interior dimensions: 1200 × 600 × 1020 mm (W × D × H)
- Exterior dimensions: 1580 × 1140 × 2005 mm (W × D × H)
- Weight, empty: 635 kg
- Voltage: 400 V, 50 Hz
- Nominal power: 11 kW
- Temperature range with humidity: 10...95 °C
- Temperature uniformity, depending on setpoint: ±0.3...±2 K
- Temperature fluctuation, depending on setpoint: ±0.1...±0.5 K
- Temperature fluctuation with humidity, depending on setpoint: ±0.1...±1 K
- Average heating-up rate between -40 °C and 180 °C: 4.8 K/min
- Cooling-down time from 180 °C to -70 °C: 120 min
- Average cooling-down time between 180 °C and -40 °C: 4 K/min
- Max. heat compensation at 25 °C and 90% RH: 800 W
- Shelves, standard / maximum: 1/11
- Load per rack: 40 kg
- Permitted load: 160 kg
- Energy consumption at 25 °C and 60% RH: 2200 Wh/h
- Wall clearance, side: 200 mm
- Wall clearance, back: 300 mm
- Sound pressure level: 69 dB(A)
- Technical Information Sheet:
- Data Sheet:
Recently viewed
You may also like
Questions about Dynamic climate chambers MKFT720-400V
What is this article?
Dynamic climate chambers MKFT720-400V is the article on this page. Luminix Health distributes Binder for laboratory use. It sits in Drying ovens & climate chambers, inside Heating, drying & cooling, in Lab instruments.
A drying oven holds a temperature above ambient to drive off water or to condition a sample. A climate chamber also holds humidity. Uniformity across the chamber, not only the set point, decides whether two samples on different shelves saw the same conditions.
Temperature equipment either holds a set point or drives a sample toward one. An oven dries. A furnace reaches the temperatures of ashing and melting. A bath surrounds a vessel with water or oil. A circulator moves that fluid. An incubator holds the mild warmth of a culture. An autoclave uses steam under pressure to sterilise. The number that matters is the temperature the article actually reaches, not the number painted on the dial. Temperature is part of the international system kept by the BIPM and compared by EURAMET.
Which category should I open next?
The category is Heating, drying & cooling. The sub-category is Drying ovens & climate chambers. The catalogue is Lab instruments.
Shelves beside this one are Muffle & tube furnaces, Heating mantles, Circulators, chillers & thermostats, Water & oil baths, Hot plates & heating blocks, Incubators, Heating & cooling (other), Temperature control accessories, Laboratory refrigerators & freezers, Autoclaves & sterilisers, Bunsen & gas burners and Heating tapes, cables & plates.
What should I check before I order?
When this page shows a safety data sheet, use that manufacturer file. Hazard text does not belong in the sales description. For a pack size, a grade or a certificate that is not on the page, send a quote request or contact the team.
How does ordering and shipping work?
Luminix Health distributes this article from the heart of Europe. Articles in stock ship within 2–5 business days. Standard delivery is free from €250 excl. VAT. Cold-chain goods and dangerous goods under ADR are never included in that free delivery. The terms are in the Shipping Policy. The full brand index is on shop by brand. Who we are is on about us.
Which other categories should I open?
The seven catalogues are Labware, Lab chemicals, Lab instruments, Lab consumables, Chromatography equipment, Lab safety equipment and Life Science & Microbiology.