Description
Constant climate chamber KBF PRO 720 with extended temperature and humidity range is a professional constant climate chamber for comprehensive material testing under controlled temperature and humidity conditions. The unit operates from a minimum temperature of -20 °C to a maximum of +100 °C and provides controllable relative humidity, enabling tests at low temperatures as well as under THB conditions at +85 °C and 85% relative humidity. Available as 1 pc, the chamber from manufacturer Binder supports material examinations of all kinds without difficulty. Optional light modules allow the KBF PRO to be configured as an ICH photostability test chamber or as a plant growth chamber, broadening its application scope in pharmaceutical and life-science laboratories. Net weight is 355 kg and gross weight is 357 kg. The extended temperature and humidity range makes the KBF PRO 720 suitable for demanding climate simulation, stability studies and reproducible environmental stress testing in research and quality-control settings where precise climate control is essential.
This article is part of Drying ovens & climate chambers, in Heating, drying & cooling.
Key features
- Pack size: 1 pc
- Model: KBF PRO 720
- Minimum operating temperature: -20 °C
- Maximum operating temperature: +100 °C
- Controllable parameter: Relative humidity
- Manufacturer: Binder
- Net weight: 355 kg
- UNSPSC: 41104900
- Interior dimensions: 1011 × 575 × 1250 mm (W × D × H)
- Exterior dimensions: 1285 × 900 × 1951 mm (W × D × H)
- Weight, empty: 312 kg
- Voltage: 208...240 V, 50/60 Hz
- Nominal power: 1.8 kW
- Temperature range: -20...100 °C
- Temperature range with humidity: 10...90 °C
- Temperature uniformity at 25 °C and 60% RH: ±0.3 K
- Temperature uniformity at 40 °C and 75% RH: ±0.3 K
- Temperature fluctuation at 25 °C and 60% RH: ±0.1 K
- Temperature fluctuation at 40 °C and 75% RH: ±0.1 K
- Recovery time after a 30 s door opening at 25 °C and 60% RH: 2 min
- Recovery time after a 30 s door opening at 40 °C and 75% RH: 5 min
- Humidity range: 10...98 % RH
- Humidity fluctuation at 25 °C and 60% RH: 1 ± % RH
- Humidity fluctuation at 40 °C and 75% RH: 1 ± % RH
- Max. heat compensation at 40 °C: 1000 W
- Inner doors: 2
- Shelves, standard / maximum: 2/16
- Load per rack: 40 kg
- Permitted load: 200 kg
- Energy consumption at 25 °C and 60% RH: 390 Wh/h
- Energy consumption at 40 °C and 75% RH: 435 Wh/h
- Energy consumption at 85 °C and 85% RH: 600 Wh/h
- Wall clearance, side: 180 mm
- Wall clearance, back: 180 mm
- Sound pressure level at 25 °C and 60% RH: 50 dB(A)
- Data Sheet:
Recently viewed
You may also like
Questions about Climate chamber KBF PRO 720 with expanded temperature / humidity range
What is this article?
Climate chamber KBF PRO 720 with expanded temperature / humidity range is the article on this page. Luminix Health distributes Binder for laboratory use. It sits in Drying ovens & climate chambers, inside Heating, drying & cooling, in Lab instruments.
A drying oven holds a temperature above ambient to drive off water or to condition a sample. A climate chamber also holds humidity. Uniformity across the chamber, not only the set point, decides whether two samples on different shelves saw the same conditions.
Temperature equipment either holds a set point or drives a sample toward one. An oven dries. A furnace reaches the temperatures of ashing and melting. A bath surrounds a vessel with water or oil. A circulator moves that fluid. An incubator holds the mild warmth of a culture. An autoclave uses steam under pressure to sterilise. The number that matters is the temperature the article actually reaches, not the number painted on the dial. Temperature is part of the international system kept by the BIPM and compared by EURAMET.
Which category should I open next?
The category is Heating, drying & cooling. The sub-category is Drying ovens & climate chambers. The catalogue is Lab instruments.
Shelves beside this one are Muffle & tube furnaces, Heating mantles, Circulators, chillers & thermostats, Water & oil baths, Hot plates & heating blocks, Incubators, Heating & cooling (other), Temperature control accessories, Laboratory refrigerators & freezers, Autoclaves & sterilisers, Bunsen & gas burners and Heating tapes, cables & plates.
What should I check before I order?
When this page shows a safety data sheet, use that manufacturer file. Hazard text does not belong in the sales description. For a pack size, a grade or a certificate that is not on the page, send a quote request or contact the team.
How does ordering and shipping work?
We deliver across the EU, usually within two to five working days. More is on Delivery across the EU. The terms are in the Shipping Policy. The full brand index is on shop by brand. Who we are is on about us.
Which other categories should I open?
The seven catalogues are Labware, Lab chemicals, Lab instruments, Lab consumables, Chromatography equipment, Lab safety equipment and Life Science & Microbiology.