Description
The Standard incubator BD720-230V is a large-capacity free-convection incubator from the BD Avantgarde.Line series for gentle incubation of organisms and conditioning of heat-sensitive media. Working temperatures run from +5 °C above ambient to +100 °C, with high accuracy provided by APT.line™ technology and homogeneous temperature distribution for consistent outcomes. Nominal volume is 737 l, and energy use is stated as up to 30% lower than typical comparable instruments. An LCD controller with program memory, electromechanical exhaust-air flap actuation, Ethernet connectivity and a USB port for data recording support structured laboratory processes. Protective and practical elements include an inner door of tempered safety glass (ESG), an integrated temperature selector monitor class 3.1 (DIN 12880) with optical alarm, two chrome-plated shelves and an ergonomic handle design. Delivered as one unit with Euro plug (type C), this incubator from manufacturer Binder is suited to high-volume incubation and media work requiring stable natural convection and clear programmable control.
This article is part of Incubators, in Heating, drying & cooling.
Key features
- Pack size: 1 pc
- Nominal volume: 737 l
- Temperature range: +5 °C above ambient to +100 °C
- Technology: APT.line™ free convection
- Controller: LCD with program memory
- Safety monitor: Class 3.1 (DIN 12880) with optical alarm
- Manufacturer: Binder
- UNSPSC: 41104401
- Interior dimensions: 960 × 600 × 1280 mm (W × D × H)
- Exterior dimensions: 1165 × 870 × 1590 mm (W × D × H)
- Weight, empty: 169 kg
- Voltage: 230 V, 50/60 Hz
- Nominal power: 1.65 kW
- Temperature uniformity at 37 °C: ±0.7 K
- Temperature uniformity at 100 °C: ±3 K
- Temperature fluctuation at 37 °C: ±0.1 K
- Temperature fluctuation at 100 °C: ±0.3 K
- Heating-up time to 100 °C: 90 min
- Heating-up time to 37 °C: 70 min
- Recovery time after a 30 s door opening at 37 °C: 23 min
- Inner doors: 2
- Shelves, standard / maximum: 2/16
- Load per rack: 45 kg
- Permitted load: 315 kg
- Energy consumption at 100 °C: 480 Wh/h
- Energy consumption at 37 °C: 78 Wh/h
- Wall clearance, side: 100 mm
- Wall clearance, back: 160 mm
- Technical Information Sheet:
- Data Sheet:
Recently viewed
You may also like
Browse related categories
Category
Sub-category
Catalogue
Brand
Questions about Standard-Incubators BD720-230V
What is this article?
Standard-Incubators BD720-230V is the article on this page. Luminix Health distributes Binder for laboratory use. It sits in Incubators, inside Heating, drying & cooling, in Lab instruments.
An incubator holds the mild, even temperature a culture needs, often 25 °C or 37 °C, sometimes with carbon dioxide. It is not a drying oven. Uniformity and, for cells, the gas control are the specification. A flask of cells in a drying oven is a dead flask.
Temperature equipment either holds a set point or drives a sample toward one. An oven dries. A furnace reaches the temperatures of ashing and melting. A bath surrounds a vessel with water or oil. A circulator moves that fluid. An incubator holds the mild warmth of a culture. An autoclave uses steam under pressure to sterilise. The number that matters is the temperature the article actually reaches, not the number painted on the dial. Temperature is part of the international system kept by the BIPM and compared by EURAMET.
Which category should I open next?
The category is Heating, drying & cooling. The sub-category is Incubators. The catalogue is Lab instruments.
Shelves beside this one are Muffle & tube furnaces, Heating mantles, Circulators, chillers & thermostats, Drying ovens & climate chambers, Water & oil baths, Hot plates & heating blocks, Heating & cooling (other), Temperature control accessories, Laboratory refrigerators & freezers, Autoclaves & sterilisers, Bunsen & gas burners and Heating tapes, cables & plates.
What should I check before I order?
When this page shows a safety data sheet, use that manufacturer file. Hazard text does not belong in the sales description. For a pack size, a grade or a certificate that is not on the page, send a quote request or contact the team.
How does ordering and shipping work?
We deliver across the EU, usually within 2 to 5 working days. More is on Delivery across the EU. The terms are in the Shipping Policy. The full brand index is on shop by brand. Who we are is on about us.
Which other categories should I open?
The seven catalogues are Labware, Lab chemicals, Lab instruments, Lab consumables, Chromatography equipment, Lab safety equipment and Life Science & Microbiology.